What is a Spectrophotometer?
A spectrophotometer is an instrument that measures how much light a material reflects or transmits at each wavelength.
The CIE defines it as an instrument for measuring the quotient of two values of a radiometric quantity at the same wavelength. NIST defines spectrophotometry as the quantitative measurement of the reflectance and transmittance of optical radiation. The regions are ultraviolet, visible and infrared.
How it works
The instrument compares the radiant flux that leaves a sample with the flux that falls on it. It repeats the measurement at each wavelength. The ordered readings are a spectrum (NIST SP 260-140).
A wavelength dispersion component, such as a grating, selects the wavelength. The spectral slit width measures how close the light is to a single wavelength. It depends on the grating ruling density, the focal length and the slit widths. The detector is commonly a photomultiplier tube (NIST SP 260-140).
NIST describes two ways to measure the blank and the sample:
- A single-beam instrument measures them one after the other. It divides the two spectra to get the transmittance.
- A double-beam instrument measures them almost at the same time, in two beams.
Stray light is detected light that is wrong for the measurement. It may have the wrong wavelength, or it may not have passed through the sample.
Transmittance and absorbance
Transmittance is the ratio of the radiant flux that passes through a sample to the flux that falls on it, at one wavelength. Absorbance is the negative logarithm of the transmittance. For an air blank, NIST SP 260-140 uses absorbance and transmittance density as the same quantity.
Measuring reflectance
For a solid surface, the instrument measures reflectance. Reflectance is the ratio of reflected to incident radiant flux (CIE e-ILV 17-24-064). The result depends on the geometrical and spectral conditions of the measurement. Two results are comparable only when the conditions are the same.
NIST took part in a comparison of 0/45 reflectance factor scales with PTB from 330 nm to 1100 nm. Its Reference Integrating Sphere is the national reference for directional-hemispherical reflectance.
Calibration
NIST maintains the national scales for reflectance and transmittance from 250 nm to 2500 nm. It also offers calibrations of spectral reflectance and transmittance. NIST Standard Reference Materials can calibrate, validate or characterize a spectrophotometer.
Two scales need a reference:
- The wavelength scale. SRM 2034 is a holmium oxide solution for ultraviolet and visible instruments with a spectral bandwidth of 3 nm or less. Its certified bands cover 240 nm to 650 nm.
- The absorbance scale. SRM 931h is a set of liquid absorbance standards. The certificate limits the spectral bandpass at four wavelengths.
A wavelength calibration error changes the measured transmittance density. NIST therefore certifies its neutral density glass references at local extrema of the spectrum, where this effect is smaller. A large instrumental bandwidth changes the apparent absorbance, so the certificates set a maximum bandwidth (NIST SP 260-140, sections 1.2 and 4.2.2).
Temperature affects the absorbance of SRM 931h. The certificate says to avoid temperatures above 25 degrees C. Read the certificate of each reference material for its limits.
Spectrophotometer, photometer, spectroradiometer and colorimeter
These four instruments answer different questions. The CIE gives each a short definition:
- A spectrophotometer measures the quotient of two values of a radiometric quantity at the same wavelength.
- A photometer measures photometric quantities. Photometry weights the light by the eye.
- A spectroradiometer measures radiometric quantities in narrow wavelength intervals.
- A colorimeter measures colorimetric quantities, such as the tristimulus values of a colour stimulus.
Airfield work
Airfield standards set limits for the reflectance of marking paint. FAA AC 150/5370-10H, Item P-620, sets a daylight directional reflectance for epoxy and methacrylate paint. White must reach at least 75% for epoxy and at least 80% for methacrylate. Yellow must reach at least 55% for both. The values are relative to magnesium oxide. It names the test method ASTM E2302.
The ASTM listing of that method describes a portable reflectometer. The paragraph does not name a spectrophotometer.
In the EU, CS-ADR-DSN sets minimum luminance factors for freshly coloured marking surfaces. The minimum is 0.75 for white and 0.45 for yellow. The standard conditions are illumination at 45 degrees, a view perpendicular to the surface and CIE standard illuminant D65 (CS ADR-DSN.U.935). The specification does not name an instrument.
CS-ADR-DSN also asks for a colour measurement of ground lights at five points inside the innermost isocandela curve (CS ADR-DSN.U.930(e)(1)). A light is a source, so this is a measurement of emitted light. The specification does not name the instrument for that measurement either.
Frequently Asked Questions
- What is a spectrophotometer?
- A spectrophotometer is an instrument that measures the reflectance or the transmittance of a material at each wavelength. The CIE defines it as an instrument for measuring the quotient of two values of a radiometric quantity at the same wavelength.
- What is spectrophotometry?
- Spectrophotometry is the quantitative measurement of the reflectance and the transmittance of optical radiation. NIST lists the ultraviolet, visible and infrared regions. The quantities include reflectance, transmittance, absorbance, emittance and fluorescence.
- What is a spectrophotometer used for?
- A spectrophotometer measures how a sample reflects or transmits light at each wavelength. The result is a spectrum. NIST uses spectrophotometry to maintain the national reflectance and transmittance scales. On airfields, standards set reflectance limits for marking paint.
- How do you calibrate a spectrophotometer?
- Check the wavelength scale and the absorbance scale against certified reference materials. NIST SRM 2034 checks the wavelength scale of ultraviolet and visible instruments. NIST SRM 931h checks the absorbance scale. Read the certificate for the bandwidth and the temperature.
- What is the difference between transmittance and absorbance?
- Transmittance is the ratio of the radiant flux that passes through a sample to the flux that falls on it, at one wavelength. Absorbance is the negative logarithm of the transmittance, for an air blank.
- What is the difference between a photometer and a spectrophotometer?
- A photometer measures photometric quantities, which weight light by the response of the eye. A spectrophotometer measures the quotient of two values of a radiometric quantity at the same wavelength. The CIE defines both instruments.
- What affects the accuracy of a spectrophotometer?
- A wavelength error, a spectral bandwidth that is too large and stray light all affect a reading. NIST reference materials verify the accuracy of the absorbance and transmittance scales of the instrument.
- Is a spectrophotometer used to check airfield markings?
- Not necessarily. FAA AC 150/5370-10H sets minimum daylight reflectance values for marking paint and names the test method ASTM E2302. The paragraph does not name a spectrophotometer as the instrument.
References
- nist-spectrophotometry
- Spectrophotometry NIST
- nist-reflectance-transmittance
- Spectral reflectance and transmittance NIST
- nist-sp-260-140
- NIST Special Publication 260-140, Technical Specifications for Certification of Spectrophotometric NTRMs NIST
- nist-srm-2034
- Certificate, Standard Reference Material 2034, Holmium Oxide Solution Wavelength Standard (240 nm to 650 nm) NIST
- nist-srm-931h
- Certificate of Analysis, Standard Reference Material 931h, Liquid Absorbance Standard for Ultraviolet and Visible Spectrophotometry NIST
- cie-ilv-spectrophotometer
- CIE e-ILV entry 17-25-008, spectrophotometer CIE (International Commission on Illumination), CIE S 017:2020
- cie-ilv-reflectance
- CIE e-ILV entry 17-24-064, reflectance CIE (International Commission on Illumination), CIE S 017:2020
- cie-ilv-photometer
- CIE e-ILV entry 17-25-019, photometer CIE (International Commission on Illumination), CIE S 017:2020
- cie-ilv-spectroradiometer
- CIE e-ILV entry 17-25-007, spectroradiometer CIE (International Commission on Illumination), CIE S 017:2020
- cie-ilv-colorimeter
- CIE e-ILV entry 17-25-022, colorimeter CIE (International Commission on Illumination), CIE S 017:2020
- easa-ear-aerodromes
- Easy Access Rules for Aerodromes (Regulation (EU) No 139/2014), including CS-ADR-DSN Issue 7 EASA
- faa-ac-150-5370-10h
- AC 150/5370-10H, Standard Specifications for Construction of Airports FAA