What is a Transfer Standard?

Metrology · Calibration 8 languages
Definition
A transfer standard is a device used as an intermediary to compare measurement standards.

A transfer standard is a device that is used as an intermediary to compare measurement standards.

Terms

The International Vocabulary of Metrology (VIM, JCGM 200:2012) has no entry named “transfer standard”. Its entry 5.9 is the transfer measurement device: a device used as an intermediary to compare measurement standards. The note to that entry says that measurement standards are sometimes used as transfer devices.

The National Bureau of Standards (NBS, now NIST) uses the shorter term. NBS Special Publication 676-II says a transfer standard is a calibrated artifact or instrument standard that is used for disseminating the unit of measurement. NBS loans it to a participant, who compares it with the participant’s own standards or instruments under normal operating conditions.

A related VIM entry is the travelling measurement standard (5.8). It is a standard, sometimes of special construction, that is intended for transport between different locations. The VIM example is a portable, battery-operated caesium-133 frequency standard.

The VIM entry 5.1 gives etalon as a second English name for the measurement standard. It defines it as a realization of the definition of a given quantity, with a stated quantity value and uncertainty, used as a reference. The French VIM uses étalon for a measurement standard. It gives dispositif de transfert for the transfer device and étalon voyageur for the travelling standard.

How it differs from other standards

The VIM names most standards by what they are. A transfer standard is named by what it does.

  • A primary standard (5.4) is established with a primary reference measurement procedure, or it is an artifact chosen by convention.
  • A secondary standard (5.5) is established through calibration against a primary standard.
  • A reference standard (5.6) is designated for the calibration of other standards in a given organization or at a given location.
  • A working standard (5.7) is used routinely to calibrate or verify measuring instruments.
  • An intrinsic standard (5.10) is based on an inherent and reproducible property of a phenomenon. The Josephson voltage standard and the quantum Hall resistance standard are the VIM examples.

So a transfer standard is a role, not a level between the reference and the working standard. The VIM note says that measurement standards are sometimes used as transfer devices.

How a transfer works

NBS describes the transfer as part of a measurement assurance program. The steps are these.

  1. The national laboratory calibrates the transfer standard and assigns a value with an uncertainty.
  2. The laboratory loans or ships it to the participant.
  3. The participant measures it as an unknown item, with the participant’s own reference standards and instruments.
  4. The participant compares its result with the assigned value. The difference is an offset.
  5. The participant uses the offset to correct the values of its reference standards.

NBS says this approach has an advantage over the usual calibration route in finding systematic error in the laboratory. Either route finds errors in the values of the reference standards. Only the transfer can find errors that come from how the laboratory uses them. For electrical transfer standards, NBS assigns values that take into account their behavior before, during and after their stay in the participant’s laboratory. A standard that behaves in an unstable way can be left out of the analysis.

The participant does not have to be a national laboratory. In the flow meter example below, NBS circulated transfer standards in industry.

What makes a good transfer standard

The transfer standard must keep its value on the way. NBS says an artifact that is stable for its quantity, such as the mass of an object, can usually be shipped without a change. A transfer standard that is not easy to move goes in an environmentally controlled container. NBS adds more standards and more observations whenever the stability is in question.

The device must also suit its use. NIST asks that a reference photometer head has a limiting aperture or a flat diffuser, so that its reference plane is clear. An illuminance meter with a poor spectral match, or with only a 3-digit display, is not adequate as a reference photometer. NIST also asks the user to keep a photometer close to its calibration temperature, or to correct for the temperature.

Metrological traceability (VIM 2.41) means that a measurement result can be related to a reference. The link is a documented unbroken chain of calibrations. Each calibration adds to the measurement uncertainty. A transfer standard is one way to make a link in that chain between two places.

The VIM names the chain in two entries. A calibration hierarchy (2.40) is a sequence of calibrations from a reference to the final measuring system, where each outcome depends on the previous one. A metrological traceability chain (2.42) is the sequence of measurement standards and calibrations that relate a measurement result to a reference.

NBS says a transfer adds another echelon to the calibration chain. The uncertainty of the transfer must be assessed, because it adds a systematic error to later values. The VIM says in the same way that measurement uncertainty necessarily increases along a sequence of calibrations. VIM Note 5 adds that traceability alone does not ensure that the uncertainty is adequate for a given purpose.

Examples

Frequency

NIST describes transfer standards for frequency as devices that receive and process radio signals that carry frequency traceable to NIST. The signals include GPS. A laboratory with such a receiver does not need to ship its own oscillator to the national laboratory. Several sites can calibrate at the same time.

Transfer standardFrequency uncertainty over a 24 h measurement
HF receiver (WWV and WWVH)±5 × 10⁻⁹
LF receiver (LORAN-C and WWVB)±1 × 10⁻¹²
GPS receiver±2 × 10⁻¹³

NIST says a radio path is a poor reference for short-term stability, because the path length changes. Averaging over a long enough time recovers the performance of the source. A transfer standard of this kind therefore suits long-term measurements.

Light

NIST has issued gas-filled airway beacon type lamps as luminous intensity transfer standards. They give about 150, 700 and 1400 cd. It has also issued incandescent lamps as luminous flux transfer standards. NIST accepts photometers for calibration, and it gives requirements for photometers that serve as transfer standards.

Flow and voltage

NBS circulated turbine flow meters in industry as transfer standards. It added a flow conditioner to each meter, because flow profiles vary from laboratory to laboratory. For voltage, NBS compared its transfer standards directly with a primary reference bank. It compared that bank every month with the Josephson effect, which realizes the volt.

On the airfield

EASA asks the operator to use a measuring unit of sufficient accuracy to analyse each light on a category II or III precision approach runway. For a light fixture type test, AC 150/5345-46F requires that the photometric test equipment is calibrated, as paragraph 6 of IES LM-35 describes. The accuracy of that unit rests on its calibration chain. TarmacView does not measure absolute intensity. It reports each light relative to its neighbours, and candela values need a calibrated photometer. Read photometric sensor for how such a sensor works.

Frequently Asked Questions

What is a transfer standard?
A transfer standard is a device used as an intermediary to compare measurement standards. The VIM calls it a transfer measurement device. NBS describes it as a calibrated artifact or instrument standard that is used for disseminating the unit of measurement.
How does a laboratory use a transfer standard?
The laboratory measures the transfer standard as an unknown item, with its own reference standards and instruments. It compares the result with the value that the national laboratory assigned. The difference is the offset, and the laboratory uses it to correct the values of its reference standards.
Is a transfer standard the same as a reference standard?
No. A reference standard is designated for calibrating other standards in one organization or at one location. A transfer standard is defined by its job, which is to be the intermediary between two standards. Sometimes one standard does both jobs.
Does the VIM define the term transfer standard?
Not under that name. The VIM defines the transfer measurement device, also called a transfer device, as a device used as an intermediary to compare measurement standards. Its note says that measurement standards are sometimes used as transfer devices.
What is the difference between a transfer standard and a travelling standard?
A travelling standard is defined by its intended transport. The VIM calls it a measurement standard, sometimes of special construction, that is intended for transport between different locations. A transfer standard is defined by its role as an intermediary. One device can be both.
Why does a transfer standard add uncertainty?
It adds one more step to the calibration chain. The VIM says that measurement uncertainty necessarily increases along a sequence of calibrations. NBS says the uncertainty of a transfer must be assessed, because it adds a systematic error to later values.
What are examples of transfer standards?
A GPS receiver is a transfer standard for frequency. NIST lists a typical frequency uncertainty of 2 parts in 10^13 for it over a 24 h measurement. For light, NIST has issued standard lamps for luminous intensity and luminous flux, and it calibrates photometers that labs use as transfer standards.
What is an etalon?
An etalon is a measurement standard. The English VIM lists etalon as a second name for the term measurement standard, in entry 5.1. It is a realization of the definition of a quantity, with a stated value and uncertainty, used as a reference. The French VIM term is étalon.

This is a reference entry. TarmacView provides drone inspection services and doesn't sell the equipment or materials described in this glossary.

Get Started Need your airfield lighting verified? TarmacView inspects approach, runway and taxiway lights and the runway surface by drone, during full airport operation. The lighting protocol comes the same day.
Keep reading

Learn more

04 PAGES

Related terms

03 TERMS
Calibration Standard
glossary

Calibration Standard

A calibration standard is a measurement standard with a stated value and uncertainty. Types of standard, examples, the etalon, traceability, and what an airfield uses.

Reference Standard and Reference Material
glossary

Reference Standard and Reference Material

A reference standard is a measurement standard that calibrates other standards. Reference vs certified reference material, working standard, traceability.

Equipment Calibration: Instrument Verification and Quality Assurance
glossary

Equipment Calibration: Instrument Verification and Quality Assurance

Equipment calibration compares a measuring instrument with a measurement standard. What it means, how it differs from adjustment and verification, traceability, how often to calibrate, and which airfield instruments documents name.

Further reading

01 ARTICLES
What a Regulator Actually Accepts as Evidence
What a Regulator Actually Accepts as Evidence
blog

What a Regulator Actually Accepts as Evidence

Regulators rarely name an accepted instrument. What they accept is a dated, traceable record showing that the aerodrome's own agreed procedure was…

AUG 15 2026 8 MIN