Calibration

Aviation Regulatory compliance Metrology Quality assurance

Definition of Calibration

Calibration is the documented, methodical process of comparing the output of a measurement instrument or system to a recognized standard, typically of higher accuracy and traceability. The aim is to detect, document, and, if necessary, correct deviations to ensure the instrument under test (IUT) maintains the accuracy required for its intended application.

According to the International Vocabulary of Metrology (VIM):
“Calibration is an operation that, under specified conditions, establishes a relation between the quantity values with measurement uncertainties provided by measurement standards and corresponding indications with associated measurement uncertainties. In a second step, this information is used to establish a relation for obtaining a measurement result from an indication.”

In aviation, calibration applies to a wide spectrum of instruments and systems, ranging from pressure altimeters, airspeed indicators, and fuel flow meters, to navigation aids such as VOR and ILS transmitters. Each system must be calibrated to a traceable standard to ensure its measurements conform to international, national, and manufacturer specifications, as outlined by ICAO, EASA, and FAA regulations.

Where is Calibration Used?

Calibration is essential wherever measurement accuracy underpins safety, compliance, and operational effectiveness. In aviation, this includes:

  • Aircraft onboard sensors (altitude, speed, temperature, pressure, fuel)
  • Ground-based navigation and surveillance equipment
  • Avionics and instrumentation maintenance/calibration laboratories
  • Environmental monitoring stations at airports (weather, noise, emissions)
  • Test and measurement devices in aircraft manufacturing and overhaul

How is Calibration Performed?

Calibration is performed by comparing the instrument’s output at a series of test points against a reference standard with superior accuracy and documented traceability. If readings deviate beyond tolerances, adjustments are made. The entire process—including pre- and post-adjustment data, uncertainties, and environmental conditions—is meticulously documented.

Key Terms: Calibration, Verification, and Adjustment

TermActionPurposeWhen Used
CalibrationCompare & (if needed) adjust to standardRestore/verify accuracyScheduled, post-repair, or after drift
VerificationCheck against tolerance (no adjustment)Confirm ongoing complianceRoutine, between calibrations
AdjustmentChange settings to bring into specificationRestore accuracy after calibrationAfter failed calibration or excessive drift
  • Calibration: Comparison with a standard and adjustment if necessary to restore accuracy.
  • Verification: Pass/fail check that readings remain within tolerance, without adjustment.
  • Adjustment: Alteration of internal settings or components to restore compliance.

These distinctions are essential for regulatory compliance and maintenance of measurement integrity.

Purpose and Importance of Calibration

Measurement Accuracy & Safety

Accurate measurements are vital for safe operations, maintenance, and manufacturing. Inaccurate readings from altimeters or airspeed indicators can threaten flight safety. Calibration ensures correct information, reducing risk.

Regulatory Compliance

Aviation operations are governed by strict regulations (ICAO, ISO/IEC 17025, FAA, EASA) requiring regular, documented calibration of critical instruments and systems. Compliance is enforced through audits and is necessary for certification.

Traceability

Calibration ensures traceability—an unbroken chain of documented calibrations linking every measurement to a recognized reference (usually SI units). This is crucial for auditability, quality assurance, and legal defense.

Process Control

In manufacturing and maintenance, calibration underpins process control by ensuring all measurements—such as torque, alignment, or environmental readings—are reliable.

Support for Industry Standards

Calibration is integral to quality management systems (ISO 9001, AS9100, EASA Part 145). Regular calibration demonstrates a commitment to best practices and customer safety.

Reference Standards and Traceability

SI Units and BIPM

The International System of Units (SI) is the foundation for all measurement. The Bureau International des Poids et Mesures (BIPM) maintains the SI and international metrology efforts.

National Metrology Institutes (NMIs)

Each country designates an NMI (e.g., NIST in the USA, PTB in Germany, NPL in the UK) responsible for maintaining primary standards traceable to the SI. These institutes provide calibration services and reference artifacts.

Traceability Pyramid

LevelExample Institution/ItemDescription
SI UnitsBIPM (International)Definition and realization of measurement units
Primary StandardsNIST, PTB, NPLHighest accuracy standards, directly compared to SI
Secondary StandardsAccredited calibration labsCalibrated against primary standards, slightly less accurate
Working StandardsCompany-owned reference instrumentsCalibrated to secondary standards, used for routine calibration
End-User InstrumentsAircraft, lab, or shop devicesCalibrated using working/reference standards

Every calibration certificate must specify the traceability chain, ensuring each measurement can be traced to SI units via an NMI.

The Calibration Process: Step-by-Step

  1. Preparation: Select appropriate, traceable reference standards with higher accuracy than the device under test (TUR ≥ 4:1). Stabilize environmental conditions (temperature, humidity, pressure).
  2. Comparison: Compare the instrument’s output at multiple test points to the reference standard, recording all data.
  3. Deviation Identification: Calculate and evaluate deviations against specifications and tolerances.
  4. Adjustment: If out of tolerance, adjust the instrument per manufacturer procedures.
  5. Repeat Measurement: Confirm the instrument now meets required accuracy by repeating comparison.
  6. Documentation: Record all results, uncertainties, environmental conditions, and adjustments. Issue a calibration certificate detailing the traceability, results, and next due date.

Aviation Example:
A pitot-static tester, calibrated against a NIST-traceable standard, applies known pressures to an airspeed indicator. Readings are compared, adjusted if necessary, and results documented for aircraft records.

Types of Calibration

  • Mechanical Calibration: For torque wrenches, scales, micrometers—ensures structural and assembly precision.
  • Electrical Calibration: For voltmeters, oscilloscopes, avionics power supplies—guarantees electrical measurement accuracy.
  • Thermal Calibration: For thermometers, RTDs, thermocouples—critical for engine/environmental monitoring.
  • Pressure Calibration: For barometers, transmitters, altimeters, pitot-static systems—vital for altitude and airspeed.
  • Dimensional Calibration: For calipers, gauge blocks—prevents assembly issues and ensures compliance.
  • Flow Calibration: For fuel/air flow meters—crucial for fuel management and engine performance.
  • Specialized Calibration: For laboratory pipettes, radio-frequency devices, and navigation/radar systems.

Calibration vs. Verification vs. Adjustment

AspectCalibrationVerificationAdjustment
DefinitionComparison to a traceable reference, with potential adjustmentPass/fail check, no changes madePhysical/software correction after failed cal
PurposeEstablish/restore/document accuracy and traceabilityOngoing assuranceCorrect deviations
FrequencyScheduled, post-repair, or after malfunctionRoutine (often daily/weekly)As required
DocumentationFull certificate, before/after data, uncertainties, environment, traceabilitySimple checklist or recordAdjustment log, in cal record
RegulatoryRequired by ICAO, ISO, FAA, EASA, etc.Required by SOP/quality systemMust be documented if performed
ExampleAnnual calibration of a pressure altimeterPre-flight verification of airspeed indicatorRecalibrating after failed calibration

Documentation, Calibration Certificates, and Regulatory Compliance

Calibration Certificates

Must include:

  • Unique instrument identification (serial, model)
  • Reference standards and traceability
  • Calibration results (before/after adjustment)
  • Measurement uncertainties (per ISO GUM)
  • Environmental conditions
  • Calibration and next due dates
  • Traceability statement
  • Calibration method/procedure
  • Signature and accreditation of technician/lab

Regulatory Compliance

  • ISO/IEC 17025: Competence for calibration labs (required for aviation).
  • ICAO: Traceable calibration and documentation for critical systems.
  • FAA/EASA: Calibration certificates required for maintenance records and organizational approval.
  • Other: Sector-specific regulations (AS9100, GMP, CLIA, etc.).

Record Retention

Calibration records must be retained per regulatory requirements (typically 2–5 years) and be readily available for audits.

Calibration Frequency and Scheduling

Calibration intervals are set via a risk-based approach, considering:

  • Manufacturer’s recommendations
  • Usage intensity
  • Regulatory/industry standards
  • Measurement criticality
  • Environmental conditions
  • Instrument stability history
Instrument TypeTypical IntervalAviation Example
Pressure Altimeter12 months or lessAnnual calibration per FAR Part 43
Airspeed Indicator12 monthsScheduled during major inspections
Torque Wrench6–12 monthsUsed in engine/structural assembly
Weather Station Sensors6–12 monthsAirport environmental monitoring
Fuel Flow Meter12 monthsFor engine/fuel system calibration
Avionics Test Equipment6–12 monthsOscilloscopes, signal generators, etc.

Intervals may be shortened for critical or heavily used instruments, or lengthened if stability is proven and risk is documented.

Best Practices for Calibration Management

  • Use only accredited, ISO/IEC 17025-compliant calibration labs.
  • Maintain a robust calibration schedule and tracking system.
  • Ensure all calibration certificates are complete, current, and traceable.
  • Conduct regular reviews of calibration intervals based on instrument history and risk.
  • Train personnel on the importance of calibration and proper documentation.
  • Keep calibration environments controlled for temperature, humidity, and cleanliness.

Summary

Calibration is the linchpin of measurement integrity, safety, and compliance in aviation and all precision industries. By adhering to global standards, ensuring traceability, and maintaining rigorous documentation, organizations can minimize risk, ensure operational reliability, and demonstrate due diligence to regulators and customers alike.

Further Reading

For expert guidance or calibration service inquiries, contact our team or explore our calibration services .

Frequently Asked Questions

What is calibration?

Calibration is the systematic comparison of a measurement instrument to a recognized standard, with documentation and, if needed, adjustment to ensure the instrument’s readings are accurate and traceable. In aviation, calibration is critical for flight safety, regulatory compliance, and measurement reliability.

How is calibration different from verification or adjustment?

Calibration involves both comparison and possible adjustment to restore accuracy, verification checks if the instrument remains within tolerance (no changes made), and adjustment means correcting the instrument after a failed calibration or drift. Only calibration establishes traceability and certifies measurement accuracy.

Why is calibration important in aviation?

Accurate measurements are vital for safe flight operations, maintenance, and manufacturing. Calibration ensures that critical instruments—like altimeters, airspeed indicators, and navigation aids—are reliable, reducing risk and meeting stringent regulatory requirements set by organizations like ICAO, FAA, and EASA.

What should a calibration certificate include?

A calibration certificate must detail the instrument’s identification, reference standards used, calibration results (before and after adjustment), measurement uncertainties, environmental conditions, traceability statement, calibration method, date, due date, and the technician’s signature and accreditation.

How often should instruments be calibrated?

Calibration intervals depend on manufacturer recommendations, usage intensity, regulatory requirements, criticality, environmental exposure, and historical stability. Aviation instruments typically require annual or more frequent calibration, especially those critical to safety and compliance.

Ensure measurement reliability

Maintain the highest standards of safety and compliance with regular, traceable calibration of your aviation and precision instruments.

Learn more

Instrument Calibration

Instrument Calibration

Instrument calibration ensures measurement accuracy by aligning instruments with known standards. It's essential for quality assurance, regulatory compliance, a...

5 min read
Quality Assurance Calibration +5
Calibration Standard

Calibration Standard

A calibration standard is a reference with a precisely determined value, fundamental for reliable, traceable calibration of instruments in science and industry....

6 min read
Calibration Metrology +3
Field Calibration

Field Calibration

Field calibration is the process of verifying and adjusting measurement instruments directly at their point of use, ensuring accuracy and compliance in real-wor...

9 min read
Calibration Quality assurance +3