What is equipment calibration?

Calibration · Metrology · Quality Assurance 8 languages
Definition
Equipment calibration is an operation that compares the readings of a measuring instrument with values from a measurement standard, and states the uncertainty.

Equipment Calibration

Equipment calibration is an operation that compares the readings of a measuring instrument with values from a measurement standard, and states the uncertainty.

Calibration is a term of metrology, the science of measurement. A calibrated instrument gives a reading that you can relate to a known reference. An uncalibrated instrument gives a reading with no stated link to a reference.

What calibration is

The International Vocabulary of Metrology (VIM) defines calibration in two steps. First, it establishes a relation between the values of a measurement standard, with their uncertainties, and the readings of the instrument, with their uncertainties. Second, it uses this relation to obtain a measurement result from a reading.

The result may be a statement, a calibration function, a calibration diagram, a calibration curve or a calibration table. In some cases it is an additive or multiplicative correction of the reading, with its uncertainty. The VIM notes that people often see the first step alone as the calibration.

A measurement standard is a realization of the definition of a quantity, with a stated value and uncertainty, that people use as a reference. A 1 kg mass with a stated uncertainty is one example in the VIM.

Calibration, adjustment, verification and validation

These are different operations:

  • Calibration finds the relation between the instrument and the standard.
  • Adjustment changes the instrument so that it gives prescribed readings for given values. The VIM says that calibration is a prerequisite for adjustment and that the two should not be confused. It also says that people wrongly call an adjustment “self-calibration”.
  • Verification gives objective evidence that an item meets specified requirements. For example, it can show that an instrument meets the limits of its specification. The VIM says that calibration should not be confused with verification of calibration.
  • Validation is a verification where the specified requirements are adequate for an intended use.

After an adjustment, the VIM says that the instrument must usually be recalibrated. The adjustment changes the relation that the first calibration found.

Traceability

A result is traceable when a documented, unbroken chain of calibrations relates it to a reference. The VIM calls this metrological traceability. Each calibration in the chain adds to the measurement uncertainty.

The VIM calls the sequence of calibrations from a reference to the final measuring system a calibration hierarchy. The uncertainty necessarily increases along the sequence. A comparison between two measurement standards counts as a calibration in one case. The comparison must check, and if needed correct, the value and uncertainty of one of them.

How often to calibrate

No rule gives one interval for all instruments. ILAC and OIML publish a joint guide, ILAC-G24 and OIML D 10, on how to choose and review calibration intervals. It is guidance for laboratories, and it names three purposes of a periodic calibration:

  • to improve the estimate of the deviation between a reference value and the instrument value, and its uncertainty, at the time of use
  • to show the uncertainty that the instrument can reach
  • to show whether the instrument changed in a way that casts doubt on the results since the last calibration

The guide lists the factors that set the interval. They include the required uncertainty, the tendency to wear and drift, the manufacturer’s recommendation, the extent of use and the environment. They also include trend data from earlier calibrations and the quality of checks between calibrations.

For the first interval, the guide points to the manufacturer’s recommendation, the expected use, the environment and the required uncertainty. It then asks for a review with the calibration results. In the staircase method, the next interval is longer if the instrument stayed within a set share of the error limit, for example 80 %. It is shorter if the instrument was outside the limit. The guide does not recommend an interval that stays the same and that nobody reviews.

Calibration in airfield documents

The airfield documents name calibration in a few places:

  • The aerodrome maintenance programme lists the checks for each facility, and calibration is one of the types of check (EASA AMC1 ADR.OPS.C.005).
  • Staff who assess runway surface condition are trained in the use, calibration and maintenance of the friction measurement device (EASA AMC2 ADR.OPS.B.015).
  • An operator may upgrade a low runway condition code when all available means support it. These include properly operated and calibrated measuring devices (EASA AMC1 ADR.OPS.B.037).
  • On a construction project, the contractor test facility follows ASTM D3666 for equipment calibration, checks and records (FAA AC 150/5370-10H, Item C-100).
  • On the same project, nuclear density machines are calibrated to ASTM D6938 (FAA AC 150/5370-10H, Item P-152). One optional text in the item asks for a calibration within 12 months before use.
  • Flight checks of visual aids may be combined with the planned calibration flights of the non-visual aids (EASA GM1 ADR.OPS.C.015). The page on flight inspection describes these flights.

These documents do not set calibration intervals for drone inspection equipment.

Test instruments for airfield lighting

FAA AC 150/5340-26C gives the most direct example. The circuits of airfield lighting are series circuits that a constant current regulator feeds. At the maximum output of 6.6 amps, a 1 % change in current can change the light output by as much as 5 %. The same 1 % increase can cut the lamp life by 20 %. The AC says that current measurement needs accurate and calibrated equipment.

The AC makes these recommendations:

  • Calibrate portable test instruments at least once a year.
  • Have a digital multimeter and its accessories checked and calibrated every year by a certified calibration laboratory, for the range to be measured. For series circuits, this range is 2.8 to 20 amps.
  • Check the accuracy of all current measuring devices, because many constant current regulators have a non-sinusoidal output waveform.

The AC is guidance. It is mandatory only for projects that the Airport Improvement Program or Passenger Facility Charges fund.

Aircraft equipment

The query “calibration of aircraft equipment” often means the aircraft compass. A compass calibration pad is an airport facility for calibrating an aircraft compass. It is a paved area where aircraft position so that the pilot can find the deviation error (FAA AC 150/5300-13B, paragraph 6.9). The pad is not the only way to do this.

Calibration and drone inspection

An instrument must be calibrated against a standard before its reading is a value with a stated uncertainty. The same rule applies to a drone camera. TarmacView reports the output of airfield lights relative to the neighbouring units and to the commanded dimming step. Absolute intensity in candela needs a calibrated photometer.

Frequently Asked Questions

What does equipment calibration mean?
Equipment calibration is an operation that compares the readings of an instrument with the values of a measurement standard, with the uncertainty of both. The result tells you how to turn a reading into a measurement result. It does not change the instrument.
Is calibration the same as adjustment?
No. Calibration only establishes the relation between the readings of the instrument and the values of the standard. Adjustment is a separate step that changes the instrument. The VIM says that calibration is a prerequisite for adjustment, and that the two should not be confused.
Is calibration the same as verification?
No. Verification gives objective evidence that an item meets specified requirements, for example that an instrument is within the limits of its specification. The VIM says that calibration should not be confused with verification of calibration.
What is metrological traceability?
Metrological traceability is the property of a measurement result that relates it to a reference through a documented, unbroken chain of calibrations. Each calibration in the chain adds to the measurement uncertainty.
How often should equipment be calibrated?
There is no single interval for all instruments. ILAC-G24 and OIML D 10 name the factors that set it, such as the tendency to drift, the extent of use and the environment. They do not recommend an interval that nobody reviews. FAA AC 150/5340-26C recommends a calibration at least once a year for portable electrical test instruments on airfield lighting.
Is calibration part of maintenance?
At an aerodrome, yes. EASA AMC1 ADR.OPS.C.005 lists calibration as one type of check in the maintenance programme. The same text lists visual inspection, cleaning and equipment stability or alignment as other types.
Which airfield instruments need calibration?
The documents name several. EASA trains runway surface assessment staff in the calibration of the friction measurement device. FAA AC 150/5340-26C asks for calibrated current meters to test constant current regulators. FAA AC 150/5370-10H asks for calibrated nuclear density machines on construction projects. FAA AC 150/5300-13B describes a pad where pilots calibrate an aircraft compass.
Can a drone camera report absolute light intensity?
Only after calibration against a measurement standard. A camera that was not calibrated gives indications, not values with a stated uncertainty. TarmacView reports light output relative to the neighbouring units and to the commanded dimming step. Absolute intensity in candela needs a calibrated photometer.

This is a reference entry. TarmacView provides drone inspection services and doesn't sell the equipment or materials described in this glossary.

Get Started Need your airfield lighting verified? TarmacView inspects approach, runway and taxiway lights and the runway surface by drone, during full airport operation. The lighting protocol comes the same day.
Keep reading

Learn more

04 PAGES

Related terms

03 TERMS
Calibration
glossary

Calibration

Calibration is the process of comparing and adjusting measurement instruments to recognized standards, ensuring accuracy, traceability, and safety, vital in aviation and all precision industries.

Instrument Calibration
glossary

Instrument Calibration

Instrument calibration ensures measurement accuracy by aligning instruments with known standards. It's essential for quality assurance, regulatory compliance, and operational safety across industries.

Cross-Calibration
glossary

Cross-Calibration

Cross-calibration compares readings of two or more instruments that measure the same quantity. How it differs from calibration, a worked example, traceability.