What is a calibrated photometer?
A calibrated photometer is an instrument that measures light weighted to the human eye, with readings that link to a reference through documented calibration.
Definition
The CIE defines a photometer as an “instrument for measuring photometric quantities”. Photometry is the measurement of quantities that refer to radiation as evaluated with a spectral luminous efficiency, for example V(lambda) or V’(lambda). The word “calibrated” adds a second claim. The readings of the instrument relate to measurement standards, and the relation has stated uncertainties. This page covers both claims. It then covers where EASA guidance for airfield lighting calls for photometric measurement.
What it measures
A photometer measures photometric quantities. These quantities weight radiation with the response of the CIE standard photometric observer, which follows the photopic function V(lambda) or the scotopic function V’(lambda). A radiometer does not apply this weighting.
The main quantities are:
- Illuminance is the luminous flux that arrives at a surface, per unit area. The unit is the lux.
- Luminance is the luminous intensity per unit of projected area in a given direction. The unit is the candela per square metre.
- Luminous flux is the quantity that results from weighting radiant flux with the photometric observer. The unit is the lumen.
- Luminous intensity is the luminous flux per unit of solid angle in a given direction. The SI unit is the candela.
Photometer, meters and radiometer
The CIE gives separate entries to the instruments in this table.
| Instrument | What it measures |
|---|---|
| Photometer | Photometric quantities |
| Illuminance meter | Illuminance |
| Luminance meter | Luminance |
| Colorimeter | Colorimetric quantities, such as the tristimulus values of a colour stimulus |
| Radiometer | Radiometric quantities |
Radiometry is the measurement of the quantities that are associated with optical radiation. A CIE note says that it is deprecated to use the word photometry in a broader sense that covers radiometry.
How a photometer gives the intensity of a light
The isocandela diagrams in CS ADR-DSN.U.940 specify airfield light units by luminous intensity in candela. A photometer does not measure intensity directly. NIST describes the method for its standard photometers. The photometer is calibrated for illuminance responsivity in amperes per lux. It then measures the illuminance from the light source at a known distance. Square the distance in metres. Multiply it by the output current of the photometer in amperes. Divide by the illuminance responsivity. The result is the luminous intensity in candela.
The distance must be larger than the minimum distance at which the source deviates from the inverse square law by a negligible amount. A source that is large compared with the distance does not follow the law well.
What calibrated means
Calibration relates the indications of the instrument to values from measurement standards, and states the uncertainties. It is not the same as adjusting the instrument. A calibration can take the form of a table, a curve, a function or a statement. In some cases it is a correction that you apply to the indication.
Traceability is a property of the result. A result is traceable when it relates to a reference through a documented, unbroken chain of calibrations. Each calibration in the chain adds to the uncertainty. Traceability does not by itself show that the uncertainty is adequate for the task.
For the candela, the reference is the SI definition of the unit. A national metrology institute realizes it. NIST in the United States, for example, maintains the candela on a group of eight standard photometers. It also calibrates artifacts that users submit, for luminous intensity, illuminance and other quantities.
NIST calibrates a photometer for illuminance responsivity by direct comparison with three of its standard photometers. The three standard photometers first determine the illuminance from a lamp at a reference point. The photometer under test is then placed at the same point. For an illuminance meter, NIST reports the error of the reading at each illuminance level. It does not adjust the reading unless the customer asks.
What limits the accuracy of a reading
A calibration certificate does not make every reading accurate. Two properties of the instrument matter.
- Spectral match. An ideal photometer responds like the V(lambda) function. The CIE index f1’ shows how far a photometer departs from it. The eight standard photometers at NIST have f1’ values from 1.8 % to 7 %. NIST also says that illuminance meters with poor spectral match are not adequate as standard photometers.
- Reference plane. The distance in the inverse square law runs from the source to the reference plane of the head. If the head does not define this plane correctly, the reading varies with the distance to the source. Errors can be serious at close distances.
ISO/CIE 19476 defines quality indices for illuminance meters and luminance meters. It also gives the measurement procedures for the indices and the standard conditions for calibration.
How often to recalibrate
The ILAC and OIML guidance on recalibration intervals says that there is no universally applicable single best practice. The first interval rests mainly on a risk assessment. The factors include:
- the measurement uncertainty that the task needs
- the manufacturer’s recommendation
- the tendency of the instrument to wear and drift
- the extent and severity of use
- environmental conditions
- transport of the instrument and its risks
The guidance also says to keep the calibration history, because it forms the basis for later interval decisions. The user should check the instrument between calibrations. The document is guidance. It applies to laboratories and to other parties that use measuring equipment.
Photometric measurement in airfield lighting
EASA guidance says that the aerodrome operator should ensure the serviceability of the lights by conducting photometric measurements. The operator does this at appropriate intervals, as part of the maintenance programme. This applies irrespective of the runway type. The text is an acceptable means of compliance (AMC), not a rule on the instrument.
For a precision approach runway of category II or III, the same AMC gives more detail:
- The preventive maintenance includes in-field measurement of the intensity, beam spread and orientation of the lights in the approach and runway lighting.
- The operator should measure all of these lights, as far as practicable, against the figures in CS ADR-DSN.U.940.
- The measurement should use a measuring unit of sufficient accuracy to analyse the characteristics of the individual lights.
- The frequency follows from traffic density, local pollution, equipment reliability and earlier results. It should not be less than twice a year for in-pavement lights and not less than once a year for other lights.
A light with an output below 50 per cent of the specified intensity has failed. The operator also defines a maintenance level above that value, to get a warning before failure. A photometric measurement gives the reading that you compare with these levels.
EASA also states that a photometric measurement may be used to evaluate the switch-over time of a lighting power supply. During a power supply changeover, the intensity of a light can drop. The switch-over time is how long the intensity, in a given direction, takes to fall from 50 % and recover to 50 %. EASA defines it for lights at intensities of 25 % or more.
The colour of a light is a separate measurement. EASA says to measure it at five points within the innermost isocandela curve. The result is a pair of chromaticity coordinates, not a candela value.
What a drone inspection does not replace
The same AMC asks for a flight check of the approach, runway and taxiway lighting at least once a year. The check shows that the pattern is correct and that the lights operate. A flight check is a different activity from a photometric measurement.
TarmacView does not report candela from the drone, because its camera is not a calibrated photometer. It reports each light relative to the neighbouring lights and to the commanded dimming step. Absolute intensity in candela needs a calibrated photometer.
Frequently Asked Questions
- What makes a photometer calibrated?
- A photometer is calibrated when its indications have been related to measurement standards under specified conditions, with the uncertainties stated. Calibration is not the same as adjustment. A calibration can also be a correction that is applied to the indication.
- What does traceability mean for a photometer reading?
- Traceability means that the reading can be related to a reference through a documented, unbroken chain of calibrations. Each calibration in the chain adds to the measurement uncertainty. For the candela, the reference is the SI definition of the unit.
- How often should a photometer be recalibrated?
- No source that was checked gives one interval for all photometers. The ILAC and OIML guidance says that there is no universally applicable single best practice. It bases the first interval mainly on a risk assessment, and it lists the manufacturer's recommendation, drift, extent of use and environment among the factors.
- What is the difference between a photometer and a radiometer?
- A photometer measures photometric quantities, which weight radiation with the response of the human eye. A radiometer is an instrument for measuring radiometric quantities, which are the quantities associated with optical radiation. The CIE gives each instrument its own entry.
- Do airfield rules require a calibrated photometer?
- No source that was checked names the instrument. EASA says that the aerodrome operator should ensure the serviceability of lights by photometric measurements, at appropriate intervals. For a category II or III precision approach runway, it says to use a measuring unit of sufficient accuracy to analyse the individual lights.
- How does a photometer give the intensity of a light in candela?
- A photometer that is calibrated for illuminance responsivity measures the illuminance at a known distance from the light. The luminous intensity then follows from the illuminance and the distance. The distance must be large enough that the inverse square law holds for the light source.
- How is a photometer calibrated?
- At NIST, a photometer is calibrated for illuminance responsivity by direct comparison with three NIST standard photometers. For an illuminance meter, NIST reports the error of the reading at each illuminance level. It does not adjust the reading unless the customer asks.
- Can a drone give the intensity of an airfield light in candela?
- TarmacView does not report candela from the drone. Its camera is not a calibrated photometer. It reports each light relative to the neighbouring lights and to the commanded dimming step. Absolute intensity in candela needs a calibrated photometer.
References
- cie-ilv-17-25-013
- CIE International Lighting Vocabulary (CIE S 017:2020), term 17-25-013, photometry CIE
- cie-ilv-17-25-019
- CIE International Lighting Vocabulary (CIE S 017:2020), term 17-25-019, photometer CIE
- cie-ilv-17-25-020
- CIE International Lighting Vocabulary (CIE S 017:2020), term 17-25-020, illuminance meter CIE
- cie-ilv-17-25-021
- CIE International Lighting Vocabulary (CIE S 017:2020), term 17-25-021, luminance meter CIE
- cie-ilv-17-25-022
- CIE International Lighting Vocabulary (CIE S 017:2020), term 17-25-022, colorimeter CIE
- cie-ilv-17-25-006
- CIE International Lighting Vocabulary (CIE S 017:2020), terms 17-25-005 radiometry and 17-25-006 radiometer CIE
- bipm-candela
- SI base unit: candela (cd) BIPM
- jcgm-vim-calibration
- International Vocabulary of Metrology (VIM, JCGM 200), 2.39 calibration JCGM / BIPM
- jcgm-vim-traceability
- International Vocabulary of Metrology (VIM, JCGM 200), 2.41 metrological traceability JCGM / BIPM
- nist-sp-250-95
- NIST Special Publication 250-95, NIST Measurement Services: Photometric Calibrations (Zong, Nadal, Tsai, Miller) NIST
- iso-cie-19476
- ISO/CIE 19476:2014 Characterization of the performance of illuminance meters and luminance meters CIE / ISO
- ilac-g24-oiml-d10
- ILAC-G24:2022 / OIML D 10:2022, Guidelines for the determination of recalibration intervals of measuring equipment ILAC / OIML
- easa-ear-aerodromes
- Easy Access Rules for Aerodromes (Regulation (EU) No 139/2014), including CS-ADR-DSN Issue 7 EASA
This is a reference entry. TarmacView provides drone inspection services and doesn't sell the equipment or materials described in this glossary.