What is Zero Setting?
Zero setting is an adjustment of a measuring instrument so that it shows zero when the quantity it measures is zero.
Three names point to this operation. The international vocabulary of metrology (VIM, JCGM 200:2012) says zero adjustment. NIST Handbook 44 says zero-setting mechanism for the device on a scale. Instrument manuals often say set zero or zeroing. This page uses the VIM and Handbook 44 terms.
Definition
VIM defines zero adjustment as an adjustment of a measuring system that gives a null indication for a zero value of the measured quantity. It is one case of adjustment. VIM defines adjustment as a set of operations on a measuring system. The operations make it give prescribed indications for given values of the quantity. So the quantity must be zero while you set zero. On a scale, this means no load on the load receptor.
VIM lists three types of adjustment: zero adjustment, offset adjustment and span adjustment. Span adjustment is sometimes called gain adjustment.
Zero error and zero accuracy
Zero error is the measurement error at a measured quantity value of zero (VIM 4.28). A zero error does not mean that the instrument has no other error. VIM notes that zero error is not the same as the absence of measurement error.
A zero indication does not prove that the quantity is zero. VIM calls the remaining doubt null measurement uncertainty (VIM 4.29). It is the uncertainty at a null or near zero indication. It covers an interval where one does not know whether the quantity is too small to detect or the indication is only noise.
Zero setting and calibration
An adjustment is not a calibration. VIM defines calibration as an operation with two steps. The first step relates the values of measurement standards to the indications of the instrument. The second step uses this relation to get a measurement result from an indication. VIM also says that calibration is not the same as verification.
Zero setting changes what the instrument shows. Calibration compares what it shows with a reference. VIM gives the order of work in two notes:
- Calibration is a prerequisite for adjustment.
- After an adjustment, the measuring system usually needs a new calibration.
VIM has no term zero calibration. A document that uses these words can mean a zero adjustment or a calibration. Read the procedure to see which one it describes.
Zero setting on scales
NIST Handbook 44 defines a zero-setting mechanism as the means provided to attain a zero balance indication with no load on the load-receiving element. It names four types:
- Manual: a non-automatic means, used by direct operation of a control.
- Semiautomatic: an automatic means that needs a single initiation by an operator.
- Automatic: an automatic means that sets zero without an operator.
- Initial: an automatic means that sets zero at switch-on, before it is ready for use.
Handbook 44 also defines a zero-load balance. It is a correct weight indication, or a representation of zero, when there is no load on the load-receiving element.
Automatic zero tracking (AZT) is a related but separate mechanism. It is an automatic means to maintain the zero-balance indication within specified limits, without an operator. Handbook 44 notes the difference. Zero tracking maintains zero within specified limits. Zero setting sets zero within limits that depend on the scale capacity.
Zero setting and taring
Taring is a different operation. Handbook 44 defines a tare mechanism as a mechanism that determines or balances out the weight of packaging, containers, vehicles or other material. That material is not part of the net weight. Zero setting refers to the empty instrument. Taring refers to the weight of something that sits on it.
Zero setting in airfield work
Airfield work uses many measuring devices, such as runway friction testers. The EASA rules for aerodromes name calibration for these devices in two places:
- AMC2 ADR.OPS.B.015, paragraph (e), lists the use, calibration and maintenance of a runway friction measurement device, where applicable. It is part of the training for personnel who assess the runway surface condition.
- AMC1 ADR.OPS.B.037, paragraph (b), lets an aerodrome operator upgrade a runway condition code of 1 or 0. The decision is supported by all available means of assessing runway slipperiness. These include properly operated and calibrated measuring devices, if available.
The EASA document does not describe a zero-setting procedure.
Frequently Asked Questions
- What is zero setting?
- Zero setting is an adjustment of a measuring instrument so that it shows zero when the quantity it measures is zero. The international vocabulary of metrology (VIM) calls it zero adjustment. On a scale, it means that the scale shows zero with no load on the load receptor.
- Is zero setting the same as zero adjustment?
- Yes, for the operation. VIM calls the operation zero adjustment of a measuring system. NIST Handbook 44 uses the term zero-setting mechanism for the device on a scale that does it. Both give a zero indication when the quantity is zero.
- What is zero error?
- Zero error is the measurement error at a measured quantity value of zero (VIM 4.28). It is not the same as no measurement error. Even at a zero indication, null measurement uncertainty remains (VIM 4.29). It covers the case where the quantity is too small to detect or the indication is only noise.
- Is zero setting the same as calibration?
- No. Zero setting is an adjustment, and an adjustment is not a calibration. VIM says calibration is a prerequisite for adjustment. After an adjustment, the system usually needs a new calibration.
- What does zero calibration mean?
- VIM has no term zero calibration. The words usually name one of two things. Zero adjustment changes what the instrument shows at zero. A calibration compares what the instrument shows with a reference. Check which of the two a document means.
- How is zero setting different from taring?
- Zero setting brings the indication to zero with no load on a scale. Taring balances out the weight of a container, packaging or vehicle that is not part of the net weight. NIST Handbook 44 defines the two with separate terms.
- What kinds of zero-setting mechanism exist on a scale?
- NIST Handbook 44 lists four. A manual mechanism works by direct operation of a control. A semiautomatic mechanism needs one initiation by an operator. An automatic mechanism sets zero without an operator. An initial mechanism sets zero at switch-on. Automatic zero tracking is a separate mechanism. It keeps the zero indication within set limits.
- Do airfield rules describe how to set zero on a measuring device?
- No. The EASA rules for aerodromes name the calibration of runway friction measuring devices, but the text does not describe a zero-setting procedure. The rules ask for properly operated and calibrated measuring devices when an operator upgrades a runway condition code.
References
- jcgm-200-2012
- JCGM 200:2012, International vocabulary of metrology (VIM), 3rd edition Joint Committee for Guides in Metrology (BIPM)
- nist-hb44-appendix-d
- NIST Handbook 44, Appendix D: Definitions (2026 edition) NIST
- easa-easy-access-rules-aerodromes
- Easy Access Rules for Aerodromes (Regulation (EU) No 139/2014), revision March 2026 EASA